Overview
EN ISO 11979-4:2008/A1:2012 is Amendment 1 to the international/European standard for ophthalmic implants - intraocular lenses (IOLs) - Part 4: Labelling and information. Published and approved by CEN in 2012, this amendment updates labelling requirements for IOLs with specific attention to toric intraocular lenses and clarifies how UV-blocking claims must be presented. The amendment is intended for consistent, safety-oriented labelling across EU member bodies and ISO members.
Key topics and requirements
- Scope of the amendment
- Modifies EN ISO 11979-4:2008 (labelling and information for IOLs).
- Approved by CEN on 29 August 2012; national adoption required by April 2013.
- Toric IOL labelling (additions to Table 2, Clause 5)
- Item 21: For toric IOLs, the label/primary package should include sphere, cylinder and axis (in + or − cylinder notation). Optionally, the spherical equivalent (SE) and absolute cylinder power (CYL) may be given.
- Item 22: For toric IOLs, the label should show orientation of the meridian of lowest positive power or most negative power as depicted in the lens drawing.
- UV-blocking information (new Clause 9)
- Any claim of UV blocking must be accompanied by the 10% cut-off wavelength expressed in nanometres (nm).
- Administrative edits
- Removal of a normative reference (ISO 11979-9:2006) and renumbering of clause list items to reflect deletions.
Practical applications
- IOL manufacturers: Update product labels, packaging inserts and primary container markings for toric lenses to comply with the added data fields and meridian orientation depiction.
- Regulatory affairs and quality teams: Use the amendment to ensure marketing and technical documentation meet European/ISO labelling expectations for clinical use and conformity assessment.
- Notified bodies and regulators: Evaluate labelling for compliance during conformity assessments and approvals.
- Clinicians and surgical teams: Benefit from clearer toric IOL labeling that supports correct lens selection and orientation in surgical planning.
- Distributors and purchasing: Ensure package information conveys power, axis, and UV protection details necessary for inventory and safe handling.
Related standards and notes
- This amendment is part of the ISO 11979 series on intraocular lenses (ophthalmic implants). It specifically amends Part 4 (labelling). Stakeholders should refer to the full EN ISO 11979-4:2008 text and national adoptions for complete labelling tables and implementation timelines.
Keywords: intraocular lenses, ophthalmic implants, labelling, toric intraocular lenses, ISO 11979-4 amendment, UV blocking, 10% cut-off wavelength, medical device labeling, IOL manufacturers, regulatory compliance.